Assembly Replacement Procedures

Front End Control Assembly

Front End Control Assembly

Removal

1. Remove the instrument outer case. Refer to the
2. Remove the instrument top brace. Refer to the
3. Refer to
W22 will not be present in instruments with Option 503.
Instruments with Options DP2, MPB, B40, CR3, or CRP will have the Enhanced Front End
Control Assembly which will have additional cables as shown in
4. The Front End Control assembly can now be unplugged from the
Keysight N9010A EXA Service Guide
removal procedure.
Bracket
removal procedure.
Figure
15-50. Remove the ribbon cables W16, W17, and W18 and
the wire harness W22 from the Front End Control assembly (1).
motherboard by leveraging up on the ejector and lifting the board up on
the other side.
Instrument Outer Case
Top Brace and Reference
Figure
15-51.