Instrument Messages

-300 to -399, Device-Specific Errors

-300 to -399, Device-Specific Errors
Err#
Message
-300
Device-specific error
-310
System error;
-310
System error; A license
will soon expire;<feature
code> will expire in <time>
-310
System error; Error
transmitting a LAN event to
the network.
-310
System Error; Ext Source
not responding
-310
System error; Failed to
initialize the PTP clock to
current time.
-310
System error; Failed to
instantiate the PTP
ordinary clock.
-310
System error; feature
<feature code> not licensed
-310
System error; Feature
expired; <feature code>
-310
System error; License
installation
failed;<feature code>
-310
System error; License
removal failed; <feature
code>
-310
System error; No license;
<feature code> will
terminate in <time>
-310
System error; The
configured PTP hardware
driver could not be
instantiated.
Keysight N9010A EXA Service Guide
Verbose/Correction Information
An instrument error occurred and the exact problem cannot be
specifically identified. Report this error to the nearest Keysight
Technologies sales or service office.
An internal system-type error has occurred. The exact problem cannot
be specifically identified. Report this error to the nearest Keysight
Technologies sales or service office.
The indicated feature/software will expire in the specified time.
Contact Keysight Technologies to purchase continued use of this
functionality.
Communication with the network driver failed.
Signal source at given IP address is not responding / IP does not
belong to a source. Check IP address and network connection.
Failure communicating with the DMC libraries' PTP controller.
Failure in the starting up the DMC libraries' PTP controller.
The specified feature, for example "N9073A-TR2" is not licensed. The
license may have expired. You cannot use it until you get a license.
The specified feature has expired. The license is no longer valid.
The license installation of the specified feature, for example
"N9073A-TR2", has failed. You should refer to the event log in the
control panel for more details.
The license removal of the specified feature, for example
"N9073A-TR2" has failed. You should refer to the event log in the
control panel for more details.
The specified feature will stop working in the specified time due to the
license expiration You will be prompted to save results and exit.
The PTP driver failed on initialization.
91