Assembly Replacement Procedures
Front Frame Assembly
Front Frame Assembly
Removal
Make sure any connectors on the front panel are removed.
1. Remove the instrument outer case. Refer to the
If the analyzer has Option PRC, Portable Configuration, and you want to remove the bail handle
and bumpers from the front frame, refer to the
section on
2. If Option EXM, External Mixing, is installed, disconnect the semi-rigid
3. Refer to
Figure 15-87
Front Frame Assembly Removal
Keysight N9010A EXA Service Guide
removal procedure.
page
575.
cable connection (2) on the LO/IF cable.
Figure
15-87. Using the T-10 driver, remove the eight screws (1),
four on each side, to detach the Front Frame Assembly from the chassis.
Instrument Outer Case
"Bail Handle Removal (Option PRC)"
573