Keysight Technologies X Series Service Manual page 517

Exa signal analyzer, 9 khz - 3.6/7/13.6/26.5/32/44 ghz
Hide thumbs Also See for X Series:
Table of Contents

Advertisement

Assembly Replacement Procedures
RF Area (Option 532, 544)
Replacement
1. Refer to
2. Replace the two screws that attach the attenuator to the bracket. Torque
3. Position the attenuator and bracket into the chassis and attach with the
4. Replace the ribbon cable and semi-rigid cables W9, W10, W11 and W19
Keysight N9010A EXA Service Guide
Figure
15-44. Position the attenuator in the bracket so that the
ribbon connector end is "up".
to 9 inch-pounds.
two screws. Torque to 9 inch-pounds.
to the attenuator. Torque the semi-rigid cables to 10 inch-pounds.
517

Advertisement

Table of Contents
loading

Table of Contents