Fan Assembly; Removal - Keysight Technologies X Series Service Manual

Exa signal analyzer, 9 khz - 3.6/7/13.6/26.5/32/44 ghz
Hide thumbs Also See for X Series:
Table of Contents

Advertisement

Assembly Replacement Procedures

Fan Assembly

Fan Assembly

Removal

1. Remove the instrument outer case. Refer to the
2. Remove the top brace. Refer to the
3. Remove the Front Frame Assembly. Refer to the
4. Refer to
Figure 15-84
Fan Assembly Removal
5. Unplug the fan wires from the motherboard. The fan assembly can now be
568
removal procedure.
removal procedure.
removal procedure.
Figure
15-84. Remove the eight screws (1) that attach the fan
assembly to the chassis.
lifted from the chassis.
Instrument Outer Case
Top Brace and Reference Bracket
Front Frame Assembly
Keysight N9010A EXA Service Guide