Assembly Replacement Procedures
RF Area (Option 532, 544)
To gain access to any of these parts, follow these steps:
1. Remove the instrument outer case. Refer to the
2. Remove the top brace. Refer to the
3. Remove the front panel. Refer to the
4. Refer to
Figure 15-37
RF Bracket Removal
508
removal procedure.
procedure.
procedure.
Figure
15-37. Remove the chassis side, right (1) by removing the
thirteen screws (2) using the T-10 driver.
Instrument Outer Case
Front Frame Assembly
Front Frame Assembly
Keysight N9010A EXA Service Guide
removal
removal