Keysight Technologies X Series Service Manual page 146

Exa signal analyzer, 9 khz - 3.6/7/13.6/26.5/32/44 ghz
Hide thumbs Also See for X Series:
Table of Contents

Advertisement

RF Section Troubleshooting (RF/Microwave Analyzers)
RF Section Description
A13 RF Front End Assembly
This assembly is a self-contained microcircuit that is repaired at the assembly
level. See
Figure
4-1.
Figure 4-1
A13 RF Front End Assembly View from Front Panel
This assembly contains the following circuits:
— Input Low-pass filter (RF input signals < 3.6 GHz)
— Low Band Limiter
— Optional Electronic Attenuator (0-24 dB attenuation control)
— Optional Low Band Preamplifier
— RF Mixer #1 (RF input signals < 3.6 GHz)
— 1st L.O. Sub-system (3.8 GHz to 8.7 GHz)
— 1st I.F. Amplifier (5122.5 MHz)
— I.F. Band pass filter (5122.5 MHz)
— I.F. Low-pass filter (5122.5 MHz)
— Mixer #2 (RF input signals 3.6 GHz to 13.599 GHz)
— Microwave Input Amplifier
— Mixer #3 (RF input signals 13.6 GHz to 26.5 GHz)
— 2nd I.F. Amplifier (322.5 MHz)
— Microwave L.O. Sub-system (3.8 GHz to 8.7 GHz)
— L.O. Doubler
146