Assembly Replacement Procedures
Reference Assembly

Removal

(Serial Prefix ≥ MY/SG/US5648)
1. Remove the instrument outer case. Refer to the
2. Remove the four flathead screws securing the top brace to the Reference
3. Remove the rear panel. Refer to the
4. For standard instruments, refer to
Figure 15-56
Reference Board Cables - Serial Prefix > MY/SG/US5648
without Option B40, MPB, or DP2
Keysight N9010A EXA Service Guide
removal procedure.
Assembly.
W12, W14, W19, W20, and W23 from the reference assembly (1). Use the
5/16 inch wrench where necessary.
"Instrument Outer Case"
"Rear Panel"
removal procedure.
Figure
15-56. Remove cables W5, W6,