Keysight X-Series Signal Analyzers
PC6 and PC7 CPU
Replacement Procedure
Notice: This document contains references to Agilent.
Please note that Agilent's Test and Measurement business
has become Keysight Technologies. For more information,
go to www.keysight.com.
Installation Note

Advertisement

loading

Summary of Contents for Keysight Technologies X Series

  • Page 1 Keysight X-Series Signal Analyzers PC6 and PC7 CPU Replacement Procedure Notice: This document contains references to Agilent. Please note that Agilent’s Test and Measurement business has become Keysight Technologies. For more information, go to www.keysight.com. Installation Note...
  • Page 2 The information contained in this document is subject to change without notice. Keysight Technologies makes no warranty of any kind with regard to this material, including but not limited to, the implied warranties of merchantability and fitness for a particular purpose.
  • Page 3 PC6 and PC7 CPU Replacement Kit PC6 and PC7 CPU Replacement Kit Products Affected: EXA N9010A/B MXA N9020A/B PXA N9030A/B MXE N9038A UXA N9040B/N9041B PXE N9048B Serial Numbers: To Be Performed By: (X) Keysight Service Center (X) Personnel Qualified by Keysight (X) Customer Estimated Installation Time: 0.5 Hours...
  • Page 4 PC6 and PC7 CPU Replacement Kit Contents PC6 CPU Replacement Kit Quantity Description Installation Note Quad-Core High Performance CPU (without disk drive) PC7 CPU Replacement Kit Quantity Description Installation Note Dual-Core Celeron CPU (without disk drive) Tools Required — Torx Driver T-10 —...
  • Page 5: Installation Procedure

    PC6 and PC7 CPU Replacement Kit Installation Procedure Electrostatic discharge (ESD) can damage or destroy electronic components. All work on electronic assemblies should be performed at a static-safe workstation. Refer to the documentation that pertains to your instrument for information about static-safe workstations and ordering static-safe accessories.
  • Page 6 PC6 and PC7 CPU Replacement Kit 2. Refer to Figure 2. Locate and remove the existing CPU assembly from the instrument by removing the 6 rear panel screws (1). The CPU assembly can be removed from the chassis by pulling straight out the back. Use the two extractors to pull the CPU assembly out from the chassis.
  • Page 7 PC6 and PC7 CPU Replacement Kit 4. If the part number of the CPU assembly that has been removed is one of those listed in step the disk drive removed from the assembly cannot be used with the CPU assembly included in this kit.
  • Page 8: Bios Settings

    To easily navigate the BIOS Setup Utility you will want to have an external USB keyboard connected to the instrument. Then, when the initial Keysight Technologies splash screen is displayed at power-up, press the key on the keyboard specified on the initial boot screen (typically the DEL key).
  • Page 9 PC6 and PC7 CPU Replacement Kit Functional Check and Time/Date Reset 1. Turn the instrument on. 2. Verify that Operating System can be found and that the Keysight Measurement Application starts up. 3. Press System, Show System and look under Options and determine which CPU the instrument software finds.
  • Page 10 PC6 and PC7 CPU Replacement Kit Data Backup Using Alignment Wizard The Alignment Data Wizard is launched directly from the instrument application software interface, so you do not need to exit the application software before proceeding. Follow the steps below to start the wizard: 1.
  • Page 11 PC6 and PC7 CPU Replacement Kit Click Next to proceed. 7. The next screen allows you to select Backup or Restore. Select Backup as shown in Figure then click Next to proceeded. Figure 4 Selecting Backup Operation Installation Note N9020-90248...
  • Page 12 PC6 and PC7 CPU Replacement Kit 8. The content of the next screen displayed depends on the Processor assembly type. If the instrument has a PC6S or PC7S Processor Assembly, then the screen contains the selections shown in Figure The available storage location options are as follows: —...
  • Page 13 PC6 and PC7 CPU Replacement Kit 9. If the default location shown is F:\AlignmentBackups... the SD card is present. Click Next and follow the wizard's on-screen instructions to back up the calibration data. Figure 6 Alignment Data Wizard Storage Options for PC6/PC7 This completes the Backup operation.
  • Page 14: Calibration Data Backup To Sd Card

    PC6 and PC7 CPU Replacement Kit Calibration Data Backup to SD Card If you removed the SD card because the instrument did not contain an SD card when it was sent in for CPU replacement, skip this portion of the installation. To maintain backwards compatibility, the replacement CPU assembly also has a removable SD memory card that previous PC6 or PC7 CPUs have.
  • Page 15 Displayed Average Noise Level Residual Responses End of installation. For assistance, contact your nearest Keysight Technologies Sales and Service Office. To find your local Keysight office access the following URL, or if in the United States, call the following telephone number: http://www.keysight.com/find/assist...
  • Page 16 This information is subject to change without notice. © Keysight Technologies 2015-2020 Edition 1, May 2020 N9020-90248 www.keysight.com...