Instrument Messages
-200 to -299, Execution Errors
-254
Media full
-255
Directory full
-256
File name not found;
-257
File name error;
-257
File name error; Allowable
extension is .csv
-257
File name error; Allowable
extension is .png
-257
File name error; Allowable
extension is .state
-257
File name error; Invalid
file name
-257
File name error; name too
long
-258
Media protected
-260
Expression error
-261
Math error in expression
-270
Macro error
-271
Macro syntax error
-272
Macro execution error
-273
Illegal macro label
-274
Macro parameter error
-275
Macro definition too long
-276
Macro recursion error
-277
Macro redefinition\ not
allowed
110
A legal command/query could not be executed because the media
was full
A legal command or query could not be executed because media
directory was full.
A legal command or query could not be executed because the file
name was not found in the specified location.
A legal command or query could not be executed because there was
an error with the file name on the device media. For example, maybe
you tried to copy to a duplicate file name.
You are using the wrong type of file extension for the current data/file
type.
You are using the wrong type of file extension for the current data/file
type.
You are using the wrong type of file extension for the current data/file
type.
The filename, directory name, or volume label syntax is incorrect.
A legal command or query could not be executed because the media
was protected. For example, the write-protect was set
An error was found with an expression type of data element. The exact
problem cannot be specifically identified.
An expression that has legal syntax could not be executed because of
a math error. For example, maybe you are dividing by zero.
Indicates that a macro-related execution error occurred.
Indicates a syntax error within the macro definition
Indicates that a syntactically legal macro program data sequence
could not be executed due to some error in the macro definition
Indicates that the macro label defined in the *DMC command was a
legal string syntax, but could not be accepted
Indicates that the macro definition improperly used a macro
parameter placeholder
Indicates that a syntactically legal macro program data sequence
could not be executed because the string or block contents were too
long for the device to handle
Indicates that a syntactically legal macro program data sequence
could not be executed because the device found it to be recursive
Indicates that a syntactically legal macro label in the *DMC command
could not be executed because the macro label was already defined