Instrument Messages
-200 to -299, Execution Errors
-220
Parameter error
-221
Settings conflict;
-222
Data out of range;
-222
Data out of range; Invalid
list data
-222
Data out of range; Two
entries already exist at
this x-axis value.
-223
Too much data
-223
Too much data; 200 spurs
found. Additional spurs
ignored.
-224
Illegal parameter value
-224
Illegal parameter value;
<Value> invalid. Fractional
values are not allowed.
-224
Illegal parameter value;
<value> out of range.
-224
Illegal parameter value;
Exceeding the max list
length
-224
Illegal parameter value;
Gated FFT is not available
while Sweep Type is set to
Swept
-224
Illegal parameter value;
Gated LO is not available
while Sweep Type is set to
FFT
-224
Illegal parameter value;
Gated Video is not
available while Sweep Type
is set to FFT
106
A problem was found with a program data element. The exact problem
cannot be specifically identified.
There are many types of settings conflict errors. See section 3.5 for
information about these errors.
A data element was found but the instrument could not be set to that
value because it was outside the range defined for the command. A
descriptive message may be appended, such as "clipped to upper
limit"
You tried to use a trace that has a number of sweep points that is
different from the current setting of sweep points.
When entering values for limit lines, you cannot have more than two
y-axis (amplitude) values entered for a specific x-axis (frequency)
value.
A data element (of block, expression, array type, or string type) had
more data then allowed by the command, or by the available memory.
There are too many spurs for the table (the limit is 200), and any
additional spurs that are found will be ignored.
An exact data value (from a list of the allowed values) was required -
but not found. See the feature description for information about the
expected parameter values.
The seconds parameter of an LXI time may not contain a fractional
portion. For example 123456789.0 is valid while 123456789.1 is not.
The value does not fall in the valid range
The list parameters have a maximum allowed length. You are trying to
set a length longer than the maximum.
The gated FFT function is not available if you have selected the swept
type of sweep. You must be in the FFT sweep type.
The FFT sweep type moves the LO frequency in steps. So the gated LO
function is not available if you have selected FFT sweep.
The gated video function is not available if you have selected the FFT
sweep type.
Keysight N9010A EXA Service Guide