Assembly Replacement Procedures
RF Area (Option 503, 507, 513, 526)
11.Refer to
Figure 15-28
Attenuator Installation
12.Connect W35 semi-rigid coax cable between the coaxial fixed attenuator
498
Figure
15-28. Attach the coaxial fixed attenuator (1) Switch 2
port 1. Torque to 10 inch-lbs.
(1) and port 1 of Switch 1 (the lower switch). Torque to 10 inch-pounds.
Keysight N9010A EXA Service Guide