Post-Repair Procedures
Post-Repair Procedures
S/W Revision Greater Than or Equal To A.01.55 and Less Than A.02.00
1. Close the instrument application software by pressing File, Exit, Enter
2. Connect a USB mouse to one of the instrument front panel USB ports.
3. Connect a USB FLASH drive to the other front panel USB port.
4. Using the mouse double-click on the Windows Explorer link on the
5. Double-click on BackupAndRestore.exe and the window shown in
Figure 16-1
Calibration Data Backup
6. Under the "Specify Backup Folder or File Name" use the mouse to select
7. Select the USB FLASH drive and press OK.
While you can use the "Make New Folder" button in the "Browse For Folder" dialog box to create
a sub-folder on the USB FLASH drive, and you can specify a filename in the "Specify Backup
Folder or File Name" window, you can also just use the root directory on the drive and let the
instrument select a default filename for you.
8. Select "Backup" and wait for the calibration database to be backed up.
9. When the backup has completed close the Calibration Data Backup And
Keysight N9010A EXA Service Guide
using the front panel keys.
desktop and navigate to the following folder:
C:\Program Files\Agilent\SignalAnalysis\Physics
16-1
will appear.
the "..." button to open the "Browse For Folder" dialog box.
Restore window.
Figure
599