Keysight Technologies X Series Service Manual page 30

Exa signal analyzer, 9 khz - 3.6/7/13.6/26.5/32/44 ghz
Hide thumbs Also See for X Series:
Table of Contents

Advertisement

Overview
Service Equipment You Will Need
Table 1-1
Required Test Equipment
Instrument
Power Meter
RF Power Sensor
(2 required)
Microwave Power
Sensor
(2 required)
Millimeter Power Sensor
(for Options > 532)
30
Critical Specifications
Dual Channel
Instrumentation Accuracy: ± 0.5%
Power Reference Accuracy: ± 0.6%
Compatible with 8480 series power sensors
dB relative mode
Frequency Range: 100 kHz to 3.6 GHz
Amplitude Range: −30 to +20 dBm
Zero Set: ± 50 nW
Zero Drift: < ± 10 nW
Measurement Noise: < 110 nW
VSWR
100 kHz to 1 MHz: ≤ 1.20:1
1 MHz to 2 GHz: ≤ 1.10:1
2 GHz to 3 GHz: ≤ 1.30:1
Cal Factor Uncertainty (std): < 1.6%
Option H84 Cal Factor: Characterized by
d
standards lab to: ± 0.6%
Input Connector: Type-N (m)
Frequency Range: 50 MHz to 26.5 GHz
Amplitude Range: −30 to +20 dB
Zero Set: ± 50 nW
Zero Drift: < ± 10 nW
Measurement Noise: < 110 nW
Cal Factor Uncertainty: < 3.0%
VSWR: 50 MHz to 100 MHz: 1.15:1
100 MHz to 2 GHz: 1.10:1
2 GHz to 12.4 GHz: 1.15:1
12.4 GHz to 18 GHz: 1.20:1
18 GHz to 26.5 GHz: 1.25:1
Input Connector: 3.5 mm (m)
Frequency Range: 50 MHz to 50 GHz
Amplitude Range: −30 to +20 dBm
Zero Set: ± 50 nW
Zero Drift: < ± 10 nW
Measurement Noise: < 100 nW
Cal Factor Uncertainty: < 4.5%
VSWR: 50 MHz to 100 MHz: 1.15:1
100 MHz to 2 GHz: 1.10:1
2 GHz to 12.4 GHz: 1.15:1
12.4 GHz to 18 GHz: 1.20:1
18 GHz to 26.5 GHz: 1.25:1
26.5 GHz to 40 GHz: 1.30:1
40 GHz to 50 GHz: 1.50:1
Input Connector: 2.4 mm (m)
Recommended
Al ternative
Model
a
Model
N1914A
E4419A/B
N1912A
8482A
e
N8482A
(Option H84)
N8482A CFT
e
8482A
(Non-H84 sensors
will increase
measurement
uncertainty)
e
N8485A
N8485A CFT
e
8485A
N8487A
8487A
Keysight N9010A EXA Service Guide
b
Use
A, P
A, P
e
A, P
e
A, P