Assembly Replacement Procedures
Disk Drive
14.Locate the Windows Vista License Sticker that came with the replacement
Figure 15-71
Windows License Sticker Rear Panel Location
While the replacement disk drive has the Windows XP operating system loaded onto it, the
license that accompanies it in this kit is for Windows Vista. This is being done within the scope of
our license agreement with Microsoft and is necessary at this time, as the instrument
applications software is not supported or qualified with any operating system other than
Windows XP.
15.Refer to the
Keysight N9010A EXA Service Guide
hard drive and apply the sticker to the rear of the instrument as shown in
Figure
15-71, covering the existing sticker if present.
Chapter 16, "Post-Repair
specified task for replacing the A5 Disk Drive assembly.
Procedures.", and perform all of the