Functional Tests
Amplitude Accuracy at 50 MHz
Figure 17-4
Amplitude Accuracy Test Setup
Procedure
1. Zero and calibrate the power meter.
2. Configure equipment as shown in
To minimize stress on the test equipment connections, support the power sensor.
3. If the auto alignment for the analyzer has not been performed within the
4. Press Mode, Spectrum Analyzer, Mode Preset on the analyzer.
5. Set up the synthesized sweeper by pressing:
6. Set up the signal analyzer by pressing:
646
connected directly to the signal analyzer input through the adapter.
past 24 hours, press System, Alignments, Align Now, All to perform the
auto alignment routine.
CW, 50, MHz
Power Level, -4, dBm
RF (On)
Input/Output, More, Freq Ref In, External
FREQ Channel, Center Freq, 50, MHz
SPAN X Scale, 2, MHz
AMPTD Y Scale, Attenuation, Mech Atten, 10, dB
Figure
17-4, with the power splitter
Keysight N9010A EXA Service Guide