Top Brace And Reference Bracket; Removal - Keysight Technologies X Series Service Manual

Exa signal analyzer, 9 khz - 3.6/7/13.6/26.5/32/44 ghz
Hide thumbs Also See for X Series:
Table of Contents

Advertisement

Assembly Replacement Procedures

Top Brace and Reference Bracket

Top Brace and Reference Bracket

Removal

1. Remove the instrument outer case. Refer to the
2. Refer to
3. To remove the Top Brace, use the T-10 driver to remove the eight screws
4. To remove the Reference Bracket (2) (if present), use the T-10 driver to
472
removal procedure.
Figure
15-6. On analyzers with serial prefix < MY/SG/US5648,
both the Top Brace (1) and the Reference Bracket (2) will be present. On
analyzers with serial prefix ≥ MY/SG/US5648 only the Top Brace (1) will
be present.
(3) (0515-0372) attaching the Top Brace to the chassis. Remove the six
flathead screws (4) (0515-1227) attaching the Top Brace to the A7
Midplane, A14 LO Synthesizer, and A15 Front End Controller assemblies.
Remove the four flathead screws (7) (0515-1227) securing the Top Brace
to either the Reference Bracket or the A16 Reference assembly.
remove the six screws (5) (including two at the rear panel) attaching the
Reference Bracket to the instrument.
"Instrument Outer Case"
Keysight N9010A EXA Service Guide