Keysight Technologies X Series Service Manual page 35

Exa signal analyzer, 9 khz - 3.6/7/13.6/26.5/32/44 ghz
Hide thumbs Also See for X Series:
Table of Contents

Advertisement

Overview
Service Equipment You Will Need
Table 1-1
Required Test Equipment
Instrument
Filters
300 MHz Low Pass
(2 Required)
1.8 GHz Low Pass
(2 Required)
4.4 GHz Low Pass
(2 Required)
5.0 GHz Low Pass
8.0 GHz Low Pass
12.4 GHz Low Pass
Adapters
Type-N (f) to Type-N (f)
Type-N (m) to Type-N
(m)
Type-N (f) to BNC (m)
Type-N (m) to BNC (m)
Type-N (m) to BNC (f)
Type-N (m) to
3.5 mm (m)
Keysight N9010A EXA Service Guide
Critical Specifications
Cutoff Frequency: 300 MHz
Rejection at > 435 MHz: > 45 dB
VSWR: ≤ 1.5:1
BNC (m, f)
Cutoff frequency: 1.8 GHz
Rejection at 2.1 GHz to 6 GHz: > 45 dB
Insertion Loss @ 1.8 GHz: < 1 dB
VSWR: ≤ 1.50:1 DC to 1.8 GHz
Connector: SMA (f) to SMA (f)
Cutoff frequency: 4.4 GHz
Rejection at > 5.5 GHz: > 42 dB
Insertion Loss @ 1.8 GHz: < 1 dB
VSWR: ≤ 1.50:1 DC to 1.8 GHz
Connector: SMA (f) to SMA (f)
3 dB frequency: 5.25 GHz
60 dB Rejection: > 6.7 GHz
VSWR: 1.35:1
3 dB frequency: 8.4 GHz
60 dB Rejection: > 10.8 GHz
VSWR: 1.35:1
3 dB frequency: 13.0 GHz
60 dB Rejection: > 16.74 GHz
VSWR: 1.35:1
Frequency: DC to 18 GHz
VSWR: ≤ 1.13:1
Frequency: DC to 18 GHz
VSWR: ≤ 1.13:1
Frequency: DC to 1.3 GHz
VSWR: ≤ 1.13:1
Frequency: DC to 1.3 GHz
VSWR: ≤ 1.13:1
Frequency: DC to 1.3 GHz
VSWR: ≤ 1.13:1
Frequency: DC to 18 GHz
VSWR: ≤ 1.08:1
Recommended
Al ternative
Model
a
Model
Telonic
0955-0455
Berkeley
TLP 300-4AB4
RLC L-1636
0955-0491
K + L
9135-0005
ML253-4499/X
360D
U-N/NP
RLC
F-30-5000-RF
RLC
F-30-8000-RF
RLC
F-30-12.4-RF
1250-1472
1250-1475
1250-1477
1250-1473
1250-1476
1250-1743
b
Use
P
P
P
P
P
P
P
P
P, T
P, T
P
P
35