Band Lock - Keysight Technologies X Series Service Manual

Exa signal analyzer, 9 khz - 3.6/7/13.6/26.5/32/44 ghz
Hide thumbs Also See for X Series:
Table of Contents

Advertisement

Service and Diagnostics Menus
Service Key Descriptions

Band Lock

Provides the ability to tune the analyzer over as large a range as can be
accommodated by the specified harmonic number and 1st LO frequency range.
As a result, this feature can be used to check performance in the frequency
band overlap regions. By definition, multi-band sweeps are not allowed.
When band lock is in effect the Advisory Event "Band Locked: Band <x>" where
"x" is a value between 0 and 6.
When band lock is not in affect the Advisory Event "Band Locked: Band <x>" is
cleared.
Dependencies/Couplings:
1. Band Lock is only available on analyzers with frequency range options
2. The individual bands available for selection also depends upon the
388
beyond 3.6 GHz
particular frequency range option.
Off or value of -1
Turns Band Lock to Off. Analyzer will tune from band to band
normally.
Band 0 or value of 0
Locks the analyzer to Band 0.
Band 1 or value of 1
Locks the analyzer to Band 1.
Band 2 or value of 2
Locks the analyzer to Band 2.
Band 3 or value of 3
Locks the analyzer to Band 3.
Band 4 or value of 4
Locks the analyzer to Band 4.
Band 5 or value of 5
Locks the analyzer to Band 5.
Band 6 or value of 6
Locks the analyzer to Band 6.
Keysight N9010A EXA Service Guide

Advertisement

Table of Contents
loading

Table of Contents