Keysight Technologies X Series Service Manual page 31

Exa signal analyzer, 9 khz - 3.6/7/13.6/26.5/32/44 ghz
Hide thumbs Also See for X Series:
Table of Contents

Advertisement

Overview
Service Equipment You Will Need
Table 1-1
Required Test Equipment
Instrument
High Sensitivity
Millimeter Power Sensor
(for Options > 532)
Low Power Microwave
Power Sensor
(2 required)
(for Option P07 only)
Power Sensor Cable
(2 required)
Standards
Frequency Standard
50 MHz,
−25 dBm Calibrator
Attenuators
10 dB
Step Attenuator
Keysight N9010A EXA Service Guide
Critical Specifications
Frequency Range: 50 MHz to 50 GHz
Amplitude Range: −70 to −20 dBm
Zero Set: ± 20 pW
Zero Drift: < ± 4 pW
Measurement Noise: < 4 pW
Cal Factor Uncertainty: < 4.5%
VSWR: 50 MHz to 100 MHz: 1.19:1
100 MHz to 4 GHz: 1.15:1
4 GHz to 12.4 GHz: 1.20:1
12.4 GHz to 18 GHz: 1.29:1
18 GHz to 34 GHz: 1.37:1
34 GHz to 40 GHz: 1.61:1
40 GHz to 50 GHz: 1.89:1
Input Connector: 2.4 mm (m)
Frequency Range: 50 MHz to 26.5 GHz
Amplitude Range: −70 to +20 dBm
Zero Set: ± 50 pW
Zero Drift: < ± 4 pW
Measurement Noise: < 4 pW
Cal Factor Uncertainty: < 3.0%
VSWR: 50 MHz to 100 MHz: 1.15:1
100 MHz to 2 GHz: 1.10:1
2 GHz to 12.4 GHz: 1.15:1
12.4 GHz to 18 GHz: 1.20:1
18 GHz to 26.5 GHz: 1.25:1
Input Connector: 3.5 mm (m)
Frequency: 10 MHz
−10
Accuracy: < ±1 e10
Frequency Drift: < 2.5 kHz
Typical VSWR: 1.06:1
Output Power Variation: ±.004 dB
Total Harmonic Content: −45 dBc
Range: 0 to 80 dB
Accuracy:
Characterized by standards lab
Calibrated at 50 MHz
VSWR: at 50 MHz: ≤ 1.05:1
Recommended
Model
8487D
e
8485D
11730A
Symmetricom
5071A
Z5602A
Opt H51
8496G
(Option 001,
H50)
f
Al ternative
Model
a
Keysight 5061B,
5071A
8496H
(Option 001, H50)
b
Use
A, P
A, P
A, P, T
A, P
A
P
31