Replacement Procedure

Hard Disk Assembly Replacement (Option 019)

Hard Disk Assembly Replacement (Option 019)

Tools Required

Removal Procedure

Refer to
Step 1. Remove the Outer Cover as described in
page
Step 2. Disconnect the power supply cable (item 1) from the Hard Disk Drive.
Step 3. Disconnect the S-ATA cable (item 2) from the Hard Disk Drive.
Step 4. Release the clamped cables from the clamps (item 3) on the Hard Disk
Assembly and the A50 CPU Module.
Step 5. Remove the four TORX T10 screws (item 4) fastening the Hard Disk Assembly.
Step 6. Lift the Hard Disk Assembly.
Step 7. Remove the four TORX T10 screws (item 5) fastening the Hard Disk Drive.
Replacement Procedure
Step 1. Reverse the order of the removal procedure.
Set a torque to 9 in-lb for torque driver when fasten the TORX T10 screws
(item 4).
Set a torque to 4.4 in-lb for torque driver when fasten the TORX T10
screws (item 5).
58
torque screwdriver, T15, and T20
torque screwdriver, T10 (set to 4.4 in-lb, and 9 in-lb)
Figure 4-3