Instrument Messages
-300 to -399, Device-Specific Errors
-310
System error; The PTP
hardware driver reported a
configuration error.
-310
System error; The PTP
ordinary clock reported a
configuration error.
-310
System error; The Trigger
alarm delayed LAN event was
not scheduled due to an
existing pending event.
-310
System error; The Trigger
alarm delayed LAN event was
not scheduled due to a
conflict with an existing
scheduled alarm.
-310
System error; The Trigger
alarm was not scheduled due
to a conflict with an
existing scheduled alarm.
-311
Memory error
-312
PUD memory lost
-313
Calibration memory lost
-314
Save/recall memory lost
-315
Configuration memory lost
-320
Storage fault;
-321
Out of memory
-330
Self-test failed
-340
Calibration failed
-340
Calibration failed; no
signal detected; <port>
-350
Queue overflow
-360
Communication error
92
Failure in the execution of the PTP driver. The most likely cause of this
error is a mismatch between versions of the PTP driver and the LXI
middleware.
Failure in execution of the DMC libraries' PTP controller.
Delayed LAN events cannot occur too close together (within 20 ms).
Delayed LAN events cannot occur too close to a scheduled Alarm
(within 20 ms).
Alarms cannot be scheduled to happen too close together (within 20
ms).
There is a physical problem with the instrument memory, such as a
parity error.
Protected user data saved by the *PUD command has been lost.
The nonvolatile calibration data used by the *CAL? command has been
lost.
The nonvolatile data saved by the *SAV? command has been lost.
The nonvolatile configuration data saved by the instrument has been
lost.
A problem was found while using data storage. The error is not an
indication of physical damage or failure of any mass storage element.
An internal operation needed more memory than was available. Report
this error to the nearest Keysight Technologies sales or service office.
A self-test failure occurred. Report this error to the nearest Keysight
Technologies sales or service office.
The instrument requires an Align All Now. Restore the alignment by
pressing System, Alignments, Align All Now.
The calibration for the <I | I-bar | Q | Q-bar> port did not succeed
because the calibration signal was not detected. Check that the < I |
I-bar | Q | Q-bar > port is properly connected to the Cal Out port.
An error occurred that did not get put in the error queue because the
queue was full.
There was a problem with instrument remote communications. The
exact problem cannot be specifically identified.
Keysight N9010A EXA Service Guide