Post-Repair Procedures
Post-Repair Procedures
Data Restore (SQL Database)
If the instrument being repaired was using instrument software version A.01.55
or newer prior to this repair it was already using an SQL database file for its
calibration data, so it can be easily restored with the instrument Alignment
Data Wizard using the following procedure.
1. Connect a USB mouse to one of the instrument front panel USB ports.
2. Locate the USB FLASH drive that was earlier used to backup the
3. Restore the instrument calibration data by pressing: System, Alignments,
4. Follow the on-screen instruction to restore the previously saved
602
instrument calibration file and insert it into one of the instrument front
panel USB ports.
Backup or Restore Align Data... The Alignment Data Wizard window as
shown in
Figure 16-2
calibration data from the USB FLASH drive.
will appear.
Keysight N9010A EXA Service Guide