RF Section Troubleshooting (RF/Microwave Analyzers)
Troubleshooting
Connect the functioning Spectrum Analyzer and appropriate high frequency
cable and connector to A14J740. Adjust the functioning Spectrum Analyzer to
measure a signal at 6122.5 MHz at +16 dBm ± 4 dB as seen in
Figure 4-8
Measure 1st L.O.
Keysight N9010A EXA Service Guide
Figure
4-8.
155