Analog/Digital IF Troubleshooting
40 MHz BW IF Section
12.The analyzer should read 22.5 MHz at −25 dBm ± 4 dB as shown in
Figure 7-19
22.5 MHz IF Input
13.If the 22.5 MHz signal is not measuring the correct power level, see
14.If the 22.5 MHz signal is within tolerance, carefully reconnect the W41
296
7-19.
Analog I.F. Troubleshooting" on page 265
cable to A3J19.
in this service guide.
Keysight N9010A EXA Service Guide
Figure
"A2