Instrument Messages
Condition Errors 201 to 299, Signal Integrity
211
Carrier(s) incorrect
or missing
213
Freq Out of Range
213
Freq Out of Range;
External LO
213
Freq Out of Range;
System input (IF)
215
Sync Error
215
Sync Error;Midamble
sync fail
215
Sync Error;No freq ref
pilot burst
215
Sync Error;No pilot
burst
215
Sync Error;Preamble
length zero
215
Sync Error;Sync code
not found
126
E
E
E
One or more external LO frequencies are out of range.
Check that the LO frequency limits are set correctly and
check the entered measurement frequencies and
measurement mode.
E
One or more system input frequencies are out of range. If
using a frequency list, check that all entries are valid for
current measurement mode.
E
W-CDMA: Cannot sync DPCCH pilot.
Cannot synchronize measurement with DPCCH pilot for
Power Control measurement, because the pilot signal
cannot be found. Make sure DPCCH is present in the
W-CDMA uplink signal, and that the slot format and
scramble code are set correctly.
E
Failed to find the uplink slot, which caused the
synchronization with the midamble to fail.
E
The pilot burst used for frequency reference is not active.
E
There is no Pilot burst detected.
E
Burst type is "Data" or "Preamble" and the measurement
cannot find a Preamble
E
Synchronization code is not found in the measured time
slot.