RF Section Troubleshooting (RF/Microwave Analyzers)
Troubleshooting
If this power level is incorrect remove W6 at A14J200. Refer to
Figure
4-7.
Adjust the functioning spectrum analyzer to measure a signal at 4.800 GHz at
+4.5 dBm ± 2 dB. See
Figure
4-9.
Figure 4-9
4.800 GHz Signal at A14J200
If this power level is incorrect, the most probable cause is the A16 Reference
Assembly. If this level is correct and yet the signal previously measured at
A14J740 is incorrect, the most probable cause is the A14 L.O. Synthesizer
Assembly. Reconnect W4 cable to A14J740. Reconnect W6 cable to A14J200.
156
Keysight N9010A EXA Service Guide