Functional Tests
Test Equipment
Analyzer
Option
Item
532, 544
Power Sensor
P03
Low Power Sensor
Pxx other
Low Power Sensor
than P03
Miscellaneous Equipment
All
Filter, 50 MHz Low Pass
526
Power Splitter, 3.5 mm
532, 544
Power Splitter, 2.4 mm
503, 507,
Power Splitter
513
503, 507,
Termination, 50Ω
513, 526
532, 544
Termination, 50Ω
a. Maximum frequency equal to or greater than the maximum frequency of the EXA.
b. PSG model numbers: E8244A, E8254A, E8247C Option 520, E8247C Option H31 or 540, E8257D Option 520,
E8257D Option 550, E8257D Option 567, E8267D Option 544.
c. Maximum frequency equal to or greater than the maximum frequency of the preamplifier option.
Keysight N9010A EXA Service Guide
Critical Specifications
Frequency Range: 50 MHz to
a
44 GHz
Amplitude Range: −30 to +20 dBm
Frequency Range: 50 MHz to 3.6 GHz
Amplitude Range: –70 to –20 dBm
Frequency Range: 50 MHz to 44 GHz
Amplitude Range: –70 to –20 dBm
Cutoff Frequency: 50 MHz
Rejection at 65 MHz: > 40 dB
Rejection at 75 MHz: > 60 dB
Rejection at 80 MHz: > 60 dB
Frequency: 10 MHz to 26.5 GHz
Nominal Insertion Loss: 6 dB
Tracking Between Ports: < 0.25 dB
Nominal Insertion Loss: 8.5 dB
Tracking Between Ports: < 0.4 dB
Nominal Insertion Loss: 6 dB
Tracking Between Ports: < 0.25 dB
Type N (m) Connector
Frequency: 30 Hz to 26.5 GHz
2.4 mm (f) Connector
Frequency: 30 Hz to 50 GHz
Al ternate
Recommended
Keysight Model
8487A
8481D
8487D
c
0955-0306
11667B
11667C
11667A
909A Option 012
85138B
Keysight
Model
8485D
8487D
8485D
11667B
635