Analog/Digital IF Troubleshooting
40 MHz BW IF Section
11.The analyzer should read 322.5 MHz at −23 dBm ± 3 dB as shown in
Figure 7-13
A15 322.5 MHz Output
If the 322.5 MHz signal is not measuring the correct power level, refer to
Section Troubleshooting (RF/Microwave Analyzers)."
Reconnect W37 to A15J900.
Measuring at this location is for convenience. The 322.5 MHz signal is an
IMPORTANT
output at A15J900. Since the W37 cable connects from this output to the
input of the Analog I.F. assembly at A2J100, the small coaxial cable has not
been tested at the A2J100 end.
286
7-13.
Chapter 4, "RF
in this service guide.