Condition Errors 701 to 799, Operation

Condition Errors 701 to 799, Operation
Errors showing that the instrument is busy doing something.
An event with the error number shown in the table means the condition has
been detected.
When the condition is cleared, an event with the error number plus 1000 is
generated.
For example, error 711 indicates that the Waiting for Trigger condition has
been detected; error 1711 indicates that condition has been cleared.
An E in the Error or Warning column means that an error is put up on the front
panel and sent out to SCPI when this condition is detected. A W in this column
means that a Warning is put up on the front panel but nothing goes out to
SCPI. Nothing in this column means nothing is put up; status bit only.
Err#
Message
701
Calibrating
703
Setting
705
Ranging
707
Sweeping
709
Measuring
711
Waiting for Trigger
713
Waiting for ARM
715
Correcting
717
Paused
719
NMR Ready Summary
721
DC Coupled
723
Printing
725
Mass Memory Busy
727
Instrument Summary
729
Program Running
Keysight N9010A EXA Service Guide