Keysight Technologies X Series Service Manual page 270

Exa signal analyzer, 9 khz - 3.6/7/13.6/26.5/32/44 ghz
Hide thumbs Also See for X Series:
Table of Contents

Advertisement

Analog/Digital IF Troubleshooting
25 MHz BW IF Section
Verifying the 300 MHz L.O. Input Power from the A16 Reference
Assembly
1. Refer to
Figure 7-6
A2 Analog I.F. Cables
2. Connect the W14 cable to the MMCX female to SMA female connector.
3. Press Freq, 300 MHz, Span, 1 MHz, Amplitude, 10 dBm, Peak Search on
270
Figure
7-6, carefully disconnect the W14 at A2J300
Use an appropriate cable to go from the SMA connector to the RF input of
a functioning spectrum analyzer.
the functioning spectrum analyzer.
Keysight N9010A EXA Service Guide

Advertisement

Table of Contents
loading

Table of Contents