Instrument Outer Case; Standard Instrument - Keysight Technologies X Series Service Manual

Exa signal analyzer, 9 khz - 3.6/7/13.6/26.5/32/44 ghz
Hide thumbs Also See for X Series:
Table of Contents

Advertisement

Assembly Replacement Procedures

Instrument Outer Case

Instrument Outer Case
If the instrument is placed on its face during any of the following procedures, be sure to use the
front panel protective cover (MP50), a soft surface, or soft cloth to avoid damage to the front
panel, keys, or input connector.
If the analyzer has Option PRC, Portable Configuration, refer to the
page 469

Standard Instrument

Removal
1. Disconnect the instrument from ac power.
2. Refer to
3. Using the T-20 driver, remove the four screws (including washers) (3)
4. Pull the instrument cover (5) off towards the rear of the instrument.
Keysight N9010A EXA Service Guide
to remove the outer case.
Figure
15-3. Using the T-20 driver, remove the four screws (two
on each side) (1) that attach the handle strap (2) on each side of the
instrument.
(two on each side) that hold the rear feet (4) (two on each side) in place.
"Option PRC"
section on
467