Front End Control Troubleshooting
A15 Front End Control Description
The table below describes the connector location and the final destinations of
the RF signal, switch control logic or bias voltage.
Table 6-2
A15 Connectors and Destinations
A15 Connector
Designation
FEC
EFEC-Lite
N/A
N/A
N/A
N/A
N/A
N/A
N/A
N/A
N/A
N/A
N/A
J1300
N/A
J801
J103
J700
J801
J800
J302
J301
J104
J102
J300
J302
J301
J300
N/A
J802
J100
J100
238
Description
EFEC,
S-EFEC
J900
Narrowband I.F. Out
J901
Wideband I.F. Out
J902
322.5 MHz I.F. In
J903
EXT I.F. In
J1100
Aux I.F. Out
J1300
Switched I.F. Filter Out
J801
YTF Bypass Out
J700
Low Band Switch Logic Out
J800
Mechanical Attenuator Logic Out
J301
Sweep Out
J102
R.F. Front End Logic Out
J302
Preselector Tune Out
J300
YTF Bias Control Out
J802
Low Noise Bypass Switch Logic Out
J100
Motherboard Connector
Destination
To A2
To A3
From A13
From A13 to A15 (Option EXM)
To Rear Panel
Reserved for future use
To SW1 & SW3 (Option MPB with
Option 507, 513, or 526) or to SW4
(Option MPB with Option 532 or 544)
To A11
To A9 & A10
Reserved for future use
To A13
N/A (test point only)
To A12
Reserved for future use
From A8
Keysight N9010A EXA Service Guide