Functional Tests
Test Equipment
Analyzer
Option
Item
All
10 dB Fixed Attenuator
(2 required)
Preamp
30 dB Fixed Attenuator
Options
(Pxx)
Cables
All
3.5 mm (1 meter)
532, 544
2.4 mm (f) to 2.4 mm (m)
(1 meter)
All
Cable, BNC (3 required)
Signal Source
All
Synthesized Sweeper
Power Meter
All
Power Meter
All
Power Sensor
503, 507,
Power Sensor
513
526
Power Sensor
634
Critical Specifications
Frequency: 50 MHz
VSWR: < 1.2:1
Accuracy: < 0.05 dB @ 50 MHz
Frequency: 10 MHz to 26.5 GHz
VSWR: < 1.4:1
Loss: < 2.0 dB
Frequency: 10 MHz to 44 GHz
VSWR: < 1.55:1 at 44 GHz
Loss: < 6.0 dB
120 cm (48 in.) BNC cable
Frequency: 10 MHz to 44 GHz
Harmonic level: < − 40 dBc
Amplitude range: 10 to − 20 dBm
Frequency Accuracy: 0.02%
Power Reference Accuracy: ±1.2%
Compatible with power sensor
Frequency Range: 50 MHz to
3.66 GHz
Amplitude Range: −70 to –20 dBm
Frequency Range: 50 MHz to
13.66 GHz
Amplitude Range: −30 to +20 dBm
Frequency Range: 50 MHz to
26.5 GHz
Amplitude Range: −30 to +20 dBm
Recommended
Keysight Model
8493C Option 010
11708A
11500E
8120-6164
10503A
a
b
PSG
E4418B
8481D
8481A
8485A
Keysight N9010A EXA Service Guide
Al ternate
Keysight
Model
8493A
Option 010
or
8493B
Option 010
8120-4921
83630B,
83640B,
83650B
E4419B
8487D
8485D

Advertisement

Table of Contents
loading

Table of Contents