RF Section Troubleshooting (Millimeter-Wave Analyzers)
Troubleshooting
Connect the functioning Spectrum Analyzer and appropriate high frequency
cable and connector to A14J740. Adjust the functioning Spectrum Analyzer to
measure a signal at 6122.5 MHz at +16 dBm ± 4 dB as seen in
Figure 5-5
Measure 1st L.O.
Keysight N9010A EXA Service Guide
Figure
5-5.
217