Instrument Messages
-200 to -299, Execution Errors
-278
Macro header not found
-280
Program error
-281
Cannot create program
-282
Illegal program name
-283
Illegal variable name
-284
Program currently running
-285
Program syntax error
-286
Program runtime error
-290
Memory use errors
-291
Out of memory
-292
Referenced name does not
exist
-293
Referenced name already
exists
-294
Incompatible type
Keysight N9010A EXA Service Guide
Indicates that a syntactically legal macro label in the *GMC? query
could not be executed because the header was not previously defined.
There was an execution error in a down-loaded program. The exact
problem cannot be specifically identified.
Indicates that an attempt to create a program was unsuccessful. A
reason for the failure might include not enough memory.
The name used to reference a program was invalid; for example,
redefining an existing program, deleting a nonexistent program, or in
general, referencing a nonexistent program.
An attempt was made to reference a nonexistent variable in a program.
Certain operations dealing with programs may be illegal while the
program is running; for example, deleting a running program might not
be possible.
Indicates that a syntax error appears in a downloaded program. The
syntax used when parsing the downloaded program is device-specific.
Indicates that the type or structure of a memory item is inadequate