Assembly Replacement Procedures
Reference Assembly
5. For instruments with Option B40, MPB, or DP2 but without Option B85,
Figure 15-57
Reference Board Cables - Serial Prefix > MY/SG/US5648 with Option B40,
MPB, or DP2
6. Refer to
Figure 15-58
Reference Support Bracket Removal
534
B1A, or B1X, refer to
Figure
W20, W42, and W50 from the reference assembly (1). Use the 5/16 inch
wrench where necessary.
Figure
15-58. Remove the reference support bracket (2) by
removing the pan-head screw (3) (0515-0372) securing the bracket to the
Reference assembly (1).
15-57. Remove cables W5, W6, W14, W19,
Keysight N9010A EXA Service Guide