Overview
Service Equipment You Will Need
Table 1-3
10 dB Step Attenuator
Nominal Attenuation (dB)
60
70
a. TME requires that Section 3 be characterized.
40
Attenuator Section
#1
#2
(10 dB)
(20 dB)
Off
On
On
On
#3
#4
(40 dB)
(40 dB)
On
Off
On
Off
Keysight N9010A EXA Service Guide
Allowable Uncertainty (dB)
< 0.035
< 0.040