Assembly Replacement Procedures
Disk Drive
10. Install the Programmed Hard Disk Drive to the Top Shield using
Figure 15-70
Top Shield/Hard Drive Assembly
11.Connect the Hard Drive ribbon cable from the Top Shield/Hard
12.Install the Top Shield/Hard Drive/Ribbon Cable assembly to the CPU
13.Refer to
550
precoated four Machine Screws (0515-5074) as shown in
Torque screws to 9 inch-pounds.
Drive/Ribbon Cable assembly to the CPU assembly. When the ribbon
cable is fully seated the two locking tabs will grip the sides of the ribbon
cable connector.
assembly using fifteen precoated Machine Screws (0515-5074). Torque
screws to 9 inch-pounds.
Figure
15-63. Replace the CPU assembly in the instrument with 6
screws. Torque to 9 inch-pounds.
Figure
Keysight N9010A EXA Service Guide
15-70.