Analog/Digital IF Troubleshooting
40 MHz BW IF Section
Verifying the 300 MHz LO Input Power from the A16 Reference
Assembly
1. Refer to
Figure 7-16
A2 Analog IF Cables
2. Connect the W14 cable to the MMCX female to SMA female adapter. Use
3. Press Freq, 300 MHz, Span, 1 MHz, Amplitude, 10 dBm, Peak Search on
Keysight N9010A EXA Service Guide
Figure
7-16, carefully disconnect the W14 at A2J300.
an appropriate cable to go from the SMA adapter to the RF input of a
functioning spectrum analyzer.
the functioning spectrum analyzer.
289