Instrument Messages

-200 to -299, Execution Errors

-200 to -299, Execution Errors
For -221 error messages, see the previous sections.
Note that Execution Errors are divided into subclasses:
-21x – Trigger errors
-22x – Parameter error
-23x – Data corrupt or stale (invalid data)
-24x – Hardware error
-25x – Mass storage error
-26x – Expression data error
-27x – Macro error
-28x – Program error (a downloaded program-related execution error)
-29x – Memory use error
Err#
Message
-200
Execution Error
-200
Execution error; Carrier
frequency outside device's
transmit band
-200
Execution error; Invalid
GSM burst timing
-200
Execution error; Invalid IP
address
-200
Execution error; Invalid
Marker Trace.
-200
Execution Error; No peak
found.
-200
Execution error; No ranges
are defined. Activate a
range.
-200
Execution Error;
Preselector centering
failed
-200
Execution Error; Signal not
stable enough to track
104
Verbose/Correction Information
A program execution error has occurred. The exact problem cannot be
specifically identified.
The entered channel/carrier frequency is not within the range of your
current mode setup selection of standard and device.
A GSM-like burst was acquired, but its timing is not valid. Ensure the
correct Burst Type has been selected.
The IP address supplied is either not valid or does not belong to a
compatible Signal Generator. Please check the IP address and
instrument connection and try again.
Cannot place markers on the reference trace, because the reference
trace is currently turned off or has no data.
No signal peak was found within the defined parameters of the search.
There are no active ranges in the range table. You will need to activate
at least one range.
Algorithm failed to center the preselector. This maybe caused by the
signal peak being too low in amplitude. Or it could be from excessive
CW input signal, alignment error, or hardware failure.
The signal that you have selected to track is changing too much for the
function to track it properly.
Keysight N9010A EXA Service Guide