RF Section Troubleshooting (RF/Microwave Analyzers)
Troubleshooting
When the analyzer is tuned to a center frequency of 50 MHz, the Low Band
switch should have minimal loss. Press Input/Output, RF Calibrator, 50 MHz,
AMPTD, Attenuation, 10 dB on the analyzer. Measure the 50 MHz calibrator
signal on the output of A11J2 using a functioning Spectrum Analyzer. The level
should be −35 dBm ± 5 dB as shown in
Figure 4-18
50 MHz Calibrator Signal at Output of W3 Cable
If the power level is incorrect the most probable cause is the low band switch
assembly. Reconnect W3 cable.
The following Low Band path items have been verified in the RF section:
— Switch control logic from the A15 Front End Control Assembly to:
— Input Attenuator A power level check
— Input Attenuator B power level check
— Low Band Switch logic and power level check
172
50 MHz Calibrator signal power level from the A16 Reference Assembly
1st L.O. power level from the A14 L.O. Synthesizer
•Input Attenuator A
•Input Attenuator B
•Low Band Switch
•Front End Assembly
Figure
4-18.
Keysight N9010A EXA Service Guide