Boot Up and Initialization Troubleshooting
Potential Problems During Boot Process
6. Refer to
Figure 2-6
A7 Midplane Board +12D LED
If yes:
If no:
7. With the instrument turned off remove the 8 screws that attach the front
8. Without disconnecting any of the cables carefully lay the front panel
9. Refer to
The instrument does have a screen saver which can disable the display backlight after a
predefined period of time. If there is any question as to whether or not this has been set by the
user prior to the current failure, and the "Inverter Enable" voltage measures too low, press a
front panel key and see if the voltage level increases to the expected level.
Keysight N9010A EXA Service Guide
Figure
2-6, verify that the +12D VDC power supply is on.
Is the +12D backlight supply voltage LED on?
Proceed to
step
7.
Replace the A6 Power Supply assembly.
panel assembly to the instrument chassis.
assembly face down on the work surface.
Figure
2-7, verify the 3 voltage levels listed in
correct.
Table 2-1
are
65