Instrument Messages
-200 to -299, Execution Errors
-231
Data questionable
-232
Invalid format
-232
Invalid format; Map
information not loaded
-233
Invalid version
-240
Hardware error
-240
Hardware error; See details
in Windows Event Log under
SA
-241
Hardware missing
-241
Hardware missing; Internal
preamp not available at all
frequency points
-241
Hardware missing; not
available for this model
number
-241
Hardware missing; Option
not installed
-250
Mass storage error;
-250
Mass storage error; Access
denied
-250
Mass storage error; Bad
path name
-250
Mass storage error; Can
only import single trace
.csv files
-250
Mass storage error; Cannot
make
-250
Mass storage error;
Different Antenna Unit
already in use
-250
Mass storage error;
Directory not found
108
Indicates that the measurement accuracy is suspect
A data element was found but it could not be used because the data
format or the data structure was not correct.
Instrument failed to load the burst mapping information from the
selected file.
A legal data element was found but could not be used because the
version of the data is incorrect. For example, state data changes as
new instrument features are added, so old state files may not work in
an instrument with a newer version of software.
A legal program command or query could not be executed because of
a hardware error. The exact problem cannot be specifically identified.
The internal data acquisition system detected a problem at startup
and logged the details in the Windows Event Log.
The operation could not be performed because of missing hardware;
perhaps the optional hardware is not installed.
The Internal Preamp is currently turned on, but the measurement is
being performed completely or partially outside the range of the
preamp. It is recommended that the user turns preamp off to ensure
consistent results across the entire measurement.
The hardware required is not part of this model
The optional hardware is not installed.
A problem was found with the mass storage device (memory, disk
drive, etc.). The exact problem cannot be specifically identified.
Access is denied.
The specified path is invalid.
Trace files containing multiple traces can not be imported. However, if
you need to recall multiple traces you can use the Save and Recall
functions rather than the Import and Export functions.
The directory or file cannot be created.
Attempt to import Corrections file with Antenna Unit that differs from
an in-use correction.
The system cannot find the path specified.
Keysight N9010A EXA Service Guide