Instrument Messages
-221 Settings Conflict Errors
-221
Settings conflict; FFT sweep
type is not available while
in Gated Video
-221
Settings conflict; Fixed
marker adjust not available
while Marker Function is on
-221
Settings conflict; Fixed
Marker Y value is not
adjustable with Normalize On
-221
Settings conflict; Freq >
3.6 GHz unavailable while
electronic attenuator
enabled
-221
Settings conflict; Frequency
Offset is not available with
Segmented Sweep
-221
Settings conflict; Frequency
Offset not available when
Frequency Scale is Log
-221
Settings conflict; Function
unavailable with MW Presel
off
-221
Settings conflict; Gate
control is Edge for Gated
FFT
-221
Settings conflict; Gate
control must be Edge for
this Gate Source
-221
Settings conflict; Gate is
not available when Marker
Count on
-221
Settings conflict; Gate
Length is not settable in
FFT sweeps
-221
Settings conflict; Gate
Length is not settable in
FFT sweeps
-221
Settings conflict; Gate
Method is not compatible
with current Sweep Type
setting
Keysight N9010A EXA Service Guide
The FFT sweep type is not available if you have selected the gated
video function.
If a Marker Function is on for a Fixed marker, the marker's reported
value is derived from the function. Therefore, you cannot directly
set the X or Y value of a Fixed marker that has a marker function
turned on.
If Normalize is on the Amplitude scale is in dB units, so adjusting
the Y value of a Fixed marker is not possible.
The electronic attenuator does not function above 3.6 GHz. So if
you have that attenuator enabled, you cannot change the center
frequency so that frequencies above 3.6 GHz are
displayed/measured.
The frequency offset feature cannot be used when you are using
the segmented sweep capability.
The frequency offset feature cannot be used when you have
selected a log scale for the frequency axis.
You cannot center or adjust the preselector because the
Microwave Preselector is currently off
You cannot use level triggering to control the gate if you are using
the FFT gating method.
You cannot use level triggering to control the gate when you are
using the currently selected gate source.
The gate function cannot be used while you have marker count
turned on.
The sweep time for FFT sweeps is set by the calculations. So
sweep time settings cannot be adjusted.
The sweep time for FFT sweeps is set by the calculations. So
sweep time settings cannot be adjusted.
If the Gate is On and you have the FFT Sweep Type manually
selected, then the Gate Method cannot be selected.
97