RF Section Troubleshooting (Millimeter-Wave Analyzers)
Troubleshooting
A9 Input Attenuator and A10 Input Attenuator Verification
Calibrator Switch Test
On the A16 Reference assembly, disconnect semi rigid cable W3 from A16J701
and measure A16J701 with a spectrum analyzer. Expected signal is 50 MHz at
−25 dBm ± 0.5 dB (RF calibrator set to 50 MHz). If signal level is incorrect,
suspect A16 Reference assembly is faulty, or the 50 MHz calibrator amplitude
requires adjustment using the field calibration software.
To verify calibrator switch operation, connect external signal source set to
50 MHz and −25 dBm to the RF input connector of analyzer under test. Press
Input/Output, RF Calibrator, Off. If the signal level at the attenuator output is
now correct, suspect the A9 Attenuator cal switch or a faulty control signal
from the A15 Front End Controller assembly.
Attenuator Check
Set the signal source connected to the analyzer input port to 0 dBm. On the
analyzer, press Amplitude, Attenuation, and change input attenuation to 0 dB.
The measuring spectrum analyzer connected to A10 Attenuator output port
should indicate a 0 dBm level.
Change the input attenuation on the analyzer under test to 2 dB (if Option FSA
is present) or to 10 dB (if Option FSA is not present). See the chart below for
expected measurement values at the A10 Attenuator output port.
Keysight N9010A EXA Service Guide
Analyzer Under Test
Attenuator Setting (dB)
0
a
2
a
4
a
6
a
8
10
20
30
40
50
60
a. These settings only available if Option FSA is installed.
Power at A10
Input Attenuator Being Tested
Output Port (dBm)
(Options 532, 544)
0 dBm (reference)
−2
−4
−6
−8
−10
−20
−30
−40
−50
−60
Both set to through path
A9
A9
A9
A9
A10
A10
A10
A10
A10
A10
215