Keysight Technologies X Series Service Manual page 641

Exa signal analyzer, 9 khz - 3.6/7/13.6/26.5/32/44 ghz
Hide thumbs Also See for X Series:
Table of Contents

Advertisement

Functional Tests
Frequency Readout Accuracy
7. Repeat
Table 17-2
Frequency Readout Accuracy Results
Span
Center
(MHz)
Frequency (MHz)
2990
127.2
54.1
7.95
0.106
1.98
517.59
1.98
832.50
Keysight N9010A EXA Service Guide
step 4
through
of
Table 17-2
is complete.
Minimum
1505
1495.9 MHz
1505
1504.56 MHz
1505
1504.8122 MHz
1505
1504.97240 MHz
1505
1504.999630 MHz
517.58316 MHz
832.49316 MHz
step 6
until the Marker Frequency Readout column
Marker Frequency
Readout
Maximum
1514.1 MHz
1505.44 MHz
1505.1878 MHz
1505.0276 MHz
1505.000370 MHz
517.59684 MHz
832.50684 MHz
641

Advertisement

Table of Contents
loading

Table of Contents