Functional Tests
Frequency Readout Accuracy
7. Repeat
Table 17-2
Frequency Readout Accuracy Results
Span
Center
(MHz)
Frequency (MHz)
2990
127.2
54.1
7.95
0.106
1.98
517.59
1.98
832.50
Keysight N9010A EXA Service Guide
step 4
through
of
Table 17-2
is complete.
Minimum
1505
1495.9 MHz
1505
1504.56 MHz
1505
1504.8122 MHz
1505
1504.97240 MHz
1505
1504.999630 MHz
517.58316 MHz
832.49316 MHz
step 6
until the Marker Frequency Readout column
Marker Frequency
Readout
Maximum
1514.1 MHz
1505.44 MHz
1505.1878 MHz
1505.0276 MHz
1505.000370 MHz
517.59684 MHz
832.50684 MHz