Overview
Service Equipment You Will Need
Table 1-1
Required Test Equipment
Instrument
Low Noise
Signal Generator
(one PSG may be used
for multiple sources.
c
See footnote
)
RF Signal Generator
(one PSG may be used
for multiple sources.
c
See footnote
)
Function Generator
Counters
Universal Counter
Meters
Digital Multimeter
Keysight N9010A EXA Service Guide
Critical Specifications
Frequency: 50 MHz to 1.0 GHz
Harmonics: < -30 dBc ≤ +13 dBm
Spectral Purity SSB Phase Noise @ 1 GHz:
−112 dBc at 100 Hz offset
−121 dBc at 1 kHz offset
−131 dBc at 10 kHz offset
VSWR: < 1.5:1
Frequency: 300 kHz to 3.6 GHz
Frequency Resolution: 1 Hz
Harmonic level: < −40 dBc
Amplitude range: −20 dBm to +13
Amplitude resolution: 0.02
Harmonic level:
< 2 GHz, < −30 dBc
≥ 2 GHz & < 26.5 GHz, < −50 dBc
VSWR:
< 20 GHz: 1.6:1
≤ 31 GHz: 1.8:1
For Option UNR or UNX:
Harmonics: < −30 dBc ≤ +13 dBm
Spectral Purity SSB Phase Noise @ 1 GHz:
−112 dBc at 100 Hz offset
−121 dBc at 1 kHz offset
−131 dBc at 10 kHz offset
VSWR: < 1.5:1
Frequency: 10 Hz to 300 kHz
Amplitude Resolution: 0.1 mv
Harmonic Distortion: −35 dBc
Frequency: 10 MHz
Gate time: 10 to 100 seconds
Must be capable of measuring signal at
+7 dBm (0.5 Vrms)
Frequency: 3 Hz to 300 kHz
AC Accuracy: ± 0.31% of reading
Recommended
Al ternative
Model
a
Model
E8257D
8663A
(Options UNR,
UNX)
c
PSG
(Options UNR,
UNX)
E8257D
8665B
(Options UNR,
(for Freq Resp
UNX)
Phase Noise > 30
kHz test only)
c
PSG
(Options UNR,
UNX)
33250A
33120A
(required for
(Option 001)
Channel Match
(Option BBA)
test)
53132A
53131A
3458A
b
Use
A, P, T
A, P
A, P
A, P, T
A, P, T
29