Instrument Messages
-400 to -499, Query Errors
-400 to -499, Query Errors
Err#
Message
-400
Query Error
-410
Query INTERRUPTED
-420
Query UNTERMINATED
-430
Query DEADLOCKED
-440
Query UNTERMINATED after
indefinite response
90
Verbose/Correction Information
There was a problem with a query command. The exact problem
cannot be specifically identified.
Some condition caused an INTERRUPTED query to occur. For
example, a query was followed by DAB or GET before a response was
completely sent.
Some condition caused an UNTERMINATED query to occur. For
example, the device was addressed to talk and an incomplete program
message was received.
Some condition caused a DEADLOCKED query to occur. For example,
both the input buffer and the output buffer are full and the analyzer
cannot continue. The analyzer automatically discards output to correct
the deadlock.
A query was received in the same program message after a query
requesting an indefinite response was executed.
Keysight N9010A EXA Service Guide
Need help?
Do you have a question about the X Series and is the answer not in the manual?