Post-Repair Procedures
Post-Repair Procedures
Data Restore (.xml File)
If the instrument being repaired was using a software version prior to A.01.55 it
would have been using an .xml database file for its calibration data, and this
data will need to be restored using the following process.
1. Close the instrument application software by pressing File, Exit, Enter
2. Connect a USB mouse to one of the instrument front panel USB ports.
3. Locate the USB FLASH drive that was earlier used to backup the
4. Using the mouse double-click on the Windows Explorer link on the
5. Locate the following file, which was previously saved to the USB FLASH
6. Highlight the calibration data file and copy the file to the instrument
7. Navigate to the folder below and paste the file by selecting Ed it, Paste
8. Verify that the CurrentDataSet.xml file was copied to the new disk drive
9. Cycle the instrument power.
When the instrument application software starts up it will see this older format
data file and convert it to the new SQL database. Any further backing up of the
data will now be done through the instrument Alignment Data Wizard
application which can be found in the instrument application software under
System, Alignments, Backup or Restore Align Data...
Keysight N9010A EXA Service Guide
using the front panel keys.
instrument calibration file and insert it into one of the instrument front
panel USB ports.
desktop and navigate to the folder on the USB FLASH drive where you
previously saved the calibration data file.
drive:
CurrentDataSet.xml
clipboard by selecting Ed it, Copy from the Windows Explorer pull-down
menu.
from the Windows Explorer pull-down menu:
E:\AlignDataStorage
601